Heating-up chamber CALDERA 300

Application:

Heating-Up Chamber CALDERA

Caldera is device for Fluid warmer

Functionality

  • capacities 68 150 200 250 300l
  • dimensions and load examples are specified in the table with technical data
  • fast heating-up of the load due to forced air convection
  • polished stainless steel housing stainless steel interior
  • bright energy saving LED internal lighting and tempered glass of the door assure an excellent visibility of the interior
  • drawers with telescopic slides instead of regular shelves prevent falling out of the load
  • possibility to place Caldera on the stainless steel table

option

  • Safety – safe temperature range: +35°C to +42°C, temperature regulation every 1°C
  • visual and audio alarm in case set temperature is exceeded for 2°C
  • independent temperature protection over 45°C (over temperature protection) 3.1 class according to DIN 12880
  • open door alarm (the alarm goes on in case the door is opened longer than 1 minute)
  • LED display visible from 4m distance
  • optional door lock – load protection against unauthorized use
  • service settings protection against unauthorized use
  • archiving working parameters for the period of 1 year (with 15 minutes interval)

CALDERA complies with the following medical norm
PN-EN 60601-1-2:2002 EMC
Medical norm for electrical devices (it does not interrupt the work of the other medical Instruments)

Note

fluid warmers caldera are also available with shelves instead of drawers for sheets blankets and towels heating with temperature range +35…+70C

Functionality:

  • capacities: 70,150, 200, 250, 300l
  • dimensions and load examples are specified in the table with technical data
  • fast heating-up of the load due to forced air convection
  • polished stainless steel housing, stainless steel interior
  • bright, energy saving LED internal lighting and tempered glass of the door assure an excellent visibility of the interior
  • drawers with telescopic slides instead of regular shelves prevent falling out of the load
  • possibility to place Caldera on the stainless steel table
  • Safety option
  • safe temperature range: +35°C to +42°C, temperature regulation every 1,0°C
  • visual and audio alarm in case set temperature is exceeded for 2°C
  • independent temperature protection over 45°C (over temperature protection) 3.1 class according to DIN 12880
  • open door alarm (the alarm goes on in case the door is opened longer than 1 minute)
  • LED display visible from 4m distance
  • optional door lock
  • load protection against unauthorized use
  • service settings protection against unauthorized use
  • archiving working parameters for the period of 1 year (with 15 minutes interval)

CALDERA compiles with the following medical norm PN-EN 60601-1-2:2002 EMC Medical norm for electrical devices (it does not interrupt the work of the other medical Instruments).

Heating-up chamber CALDERA 300

Technical data Smart

Informazioni aggiuntive

type

CALDERA 300

air convection

forced

chamber capacity (l)

300

working capacity (l)

300

controller

microprocessor with external LED display

display

LED

temperature range (°C)

+35…+42

temperature resolution every (°C)

1

temperature fluctuation at 37°C (+/-°C)

0,3

temperature variation at 37°C (+/-°C)

0,5

over temperature protection

class 3.1 to DIN 12880

door

door with viewing window

interior

acid-proof stainless steel to DIN 1.4301

housing

stainless steel polished

overall dims (mm) /1/

_bw

A width

600

B height

1440

depth C

630

internal dims (mm)

_bw

width D

490

height E

1250

depth F

480

number of drawers

4

max drawer workload (kg)

20

max unit workload (kg)

80

weight (kg)

75

power supply

230V 50-60Hz

nominal power (W)

250

warranty

24 months

all the above technical data refer to standard units (without optional accessories)
* – fluctuation measured in centre of the chamber; in space, variation (K) calculated for chamber as:
K= +/- (T average max. – T average min. ) / 2
** – other power supplies on request
1 – depth doesn’t include 50 mm of power cable