Cooled Incubator (IL) ILW 750 Smart PRO

Application:

Cooled incubators are perfect for incubation of samples in a stable environment, regardless of ambient conditions, at temperatures from -10 up to +100°C.
– microbiological tests
– plant growing, microorganisms breeding at precisely controlled environment
– BOD determination
– incubation of samples at specified temperature
– Special models:
– FOT – photoperiodic system – day and night simulation
– FIT – phytotron system – day and night simulation with intensity control

Advantages of the SMART PRO controller:

  • large (7 inches), clear, full colour touch screen
  • LAN, USB ports and WiFi for communication and data transfer
  • multi-segment time and temperature programs
  • overview of data in tabular and graphic form
  • visual and sound alarm
  • Admin function for management
  • password protected log-in
  • internal memory for programs and data storage
  • event registry with user notifications
  • LabDesk software and user manual for direct download

Cooled Incubator (IL) ILW 750 Smart PRO

Technical data Smart

Informazioni aggiuntive

type

ILW 750 Smart PRO

air convection

forced

chamber capacity (l)

749

working capacity (l)

749

controller

microprocessor PID

display

7" full colour touch screen

temperature range (°C)

0…+100 / -10…+100 (option)

temperature resolution every (°C)

0,1

temperature fluctuation at 37°C (+/-°C)

0,2

temperature variation at 37°C (+/-°C)

0,3

over temperature protection

class 3.3 to DIN 12880

door

double(5) / door with viewing window (option)

interior

_bw

housing

_bw

Smart PRO

powder coated sheet

IG Smart PRO

stainless steel linen finish

overall dims (mm) /1/

_bw

A width

1260

B height

1910

depth C

890

internal dims (mm)

_bw

width D

1040

height E

1200

depth F

600

shelves (standard - max)

5, 16

reinforced shelf version (PW) (kg) /3/

100

max unit workload (kg)

140

reinforced unit version (W) (kg) /4/

300

weight (kg)

256

power supply

230V 50-60Hz

nominal power (W)

1900

refrigerant

R290 / GWP=3

warranty

24 months

all the above technical data refer to standard units (without optional accessories)
* – fluctuation measured in centre of the chamber; in space, variation (K) calculated for chamber as:
K= +/- (T average max. – T average min. ) / 2
** – other power supplies on request
1 – depth doesn’t include 50 mm of power cable
2 – on uniformly loaded surface
3 – reinforced shelf
4 – reinforced version
5 – additional internal glass door