Climatic Chamber with Phytotron System KK 500 FIT P Smart PRO

Application:

Climatic chambers with phytotron is the only group of units that allows to control humidity temperature and lighting altogether
Climatic chambers with phytotron with lighting in door

  • growth of plants and fungus
  • seeds germination
  • microorganisms and insects breeding
  • photostability tests
  • food preservation tests
  • any kind of research that requires a stable temperature and humidity environment (optionally light)
  • tests of building materials

Advantages of the SMART PRO controller:

  • large (7 inches), clear, full colour touch screen
  • LAN, USB ports and WiFi for communication and data transfer
  • multi-segment time and temperature programs
  • overview of data in tabular and graphic form
  • visual and sound alarm
  • Admin function for management
  • password protected log-in
  • internal memory for programs and data storage
  • event registry with user notifications
  • LabDesk software and user manual for direct download

Climatic Chamber with Phytotron System KK 500 FIT P Smart PRO

Technical data Smart

Informazioni aggiuntive

type

KK 500 FIT P Smart PRO

air convection

forced

chamber capacity (l)

462

working capacity (l)

365

controller

microprocessor PID

display

7" full colour touch screen

temperature range (°C)

0…+60 (with light on +10…+50)

temperature resolution every (°C)

0,1

temperature variation at 25°C and 60%rH (+/-°C)

2

relative humidity range (%)

30…90 (see chart for details)

humidity resolution every (%)

1

humidity variation at 25°C and 60% rH (%rH)

5

over temperature protection

class 3.3 to DIN 12880

door

double (external solid, internal glass) / external glass (option)

interior

_bw

housing

_bw

Smart PRO

powder coated sheet

IG Smart PRO

stainless steel linen finish

overall dims (mm) /1/

_bw

A width

640

B height

1990

depth C

1010

internal dims (mm)

_bw

width D

480

width D 2

510

height E

1510

depth F

600

height I

1360

shelves (standard - max)

3, 11

max shelf workload (kg) /2/

20

reinforced shelf version (PW) (kg) /3/

100

max unit workload (kg)

100

weight (kg)

130

power supply

230V 50-60Hz

nominal power (W)

2150

refrigerant

R290 / GWP=3

warranty

24 months

all the above technical data refer to standard units (without optional accessories)
* – fluctuation measured in centre of the chamber; in space, variation (K) calculated for chamber as:
K= +/- (T average max. – T average min. ) / 2
** – other power supplies on request
1 – depth doesn’t include 50 mm of power cable
2 – on uniformly loaded surface
3 – reinforced shelf
4 – reinforced version